HDF without the need for drip infusion into the blood line: push/pull HDF
We have developed push/pull HDF which was an autocontrolled system for simultaneous HF and HD with no need for drip infusion into the blood line. The inulin clearance of this system was much more than HD and comparable to 18 L-replacement HDF (present system, 42.3±8.5ml/min; HD, 11.9±4.8ml/min; conv...
Gespeichert in:
Veröffentlicht in: | Journal of Japanese Society for Dialysis Therapy 1986/07/28, Vol.19(7), pp.751-756 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | jpn |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We have developed push/pull HDF which was an autocontrolled system for simultaneous HF and HD with no need for drip infusion into the blood line. The inulin clearance of this system was much more than HD and comparable to 18 L-replacement HDF (present system, 42.3±8.5ml/min; HD, 11.9±4.8ml/min; conventional HDF, 38.3±5.6ml/min). This system showed less OF rate drop during the treatment than HF, because of the intermittent reverse transmembrane flush in the hemofilter. The middle molecule sieving coefficient of the hemofilter membrane was found to decline markedly as HF progressed. The sieving coefficient of the hemofilter membrane for inulin (MW 5, 200) was 0.88±0.04 before use against 0.26±0.12 after use (p |
---|---|