Technical note: A mains drop-out simulator

During an investigation into mains-borne and radiated interference in the Ultrasound Department al St James's University Hospital, the need arose to determine the susceptibility of some equipment to the mains supply failing for very brief periods. This is known as 'drop-out'. A circui...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of medical engineering & technology 1988, Vol.12 (5), p.224-225
1. Verfasser: Norwood, H. M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:During an investigation into mains-borne and radiated interference in the Ultrasound Department al St James's University Hospital, the need arose to determine the susceptibility of some equipment to the mains supply failing for very brief periods. This is known as 'drop-out'. A circuit was built that can drop out the mains supply lo a piece of equipment for up to 15 consecutive half'cycles at a selectable repetition rate between approximately 0-4/s to 6.5/s. Circuit design was simplified by using a solid state relay (SSR), which turns on and off only at the zero crossing points in the mains supply waveform.