Use of a Gadolinium Filter on an X-ray Contrast Examination
The use of gadolinium (Gd) as a “band pass” filter for the X-ray contrast examination is discussed. Vinyl tubes filled with iodine solution were placed in a water phantom (15 cm thickness). When this phantom was exposed to a 70 kVp X-ray filtered by a Gd-filter, to give an equal film density as an a...
Gespeichert in:
Veröffentlicht in: | Kurume medical journal 1984/09/10, Vol.31(1), pp.15-20 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The use of gadolinium (Gd) as a “band pass” filter for the X-ray contrast examination is discussed. Vinyl tubes filled with iodine solution were placed in a water phantom (15 cm thickness). When this phantom was exposed to a 70 kVp X-ray filtered by a Gd-filter, to give an equal film density as an aluminium (Al) filter, the surface dose was reduced by 45 % . The effective attenuation coefficient of the iodine for an 80 kVp X-ray filtered with a Gdfilter of 0.4 mm thickness corresponds to that for a 70 kVp X-ray filtered with an Al-filter of 2 mm thickness. These results indicate that a decrease of skin dosage and an increase of film contrast are obtained by using a Gd-filter for the X-ray examination with an iodine contrast medium. |
---|---|
ISSN: | 0023-5679 1881-2090 |
DOI: | 10.2739/kurumemedj.31.15 |