Real-Time, In Vivo Measurement of Radiation Dose during Radioimmunotherapy in Mice Using a Miniature MOSFET Dosimeter Probe

This report presents the first real-time measurement of absorbed radiation dose during radioimmunotherapy in mice. Dose rate and total dose at the center of the tumor were measured after administration of ^{90}{\rm Y}\text{-labeled}$ antibodies using a miniature metal oxide semiconductor field-effec...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation Research 1995-03, Vol.141 (3), p.330-335
Hauptverfasser: Gladstone, David J., Chin, Lee M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This report presents the first real-time measurement of absorbed radiation dose during radioimmunotherapy in mice. Dose rate and total dose at the center of the tumor were measured after administration of ^{90}{\rm Y}\text{-labeled}$ antibodies using a miniature metal oxide semiconductor field-effect transistor radiation dosimeter probe which was inserted into the center of the tumor volume. Continuous real-time measurements were made for as long as 23 h after injection of the radiolabeled antibodies. Comparison of the real-time dose-rate measurements with estimates based on the MIRD formalism indicates good agreement. The real-time measurements are further compared to measurements made in a second experiment in which groups of mice were sacrificed at individual times after injection of the same radiolabeled antibodies. The real-time measurements agree well with the measurements in excised tumors. The real-time measurements have greater time resolution and are much more efficient than traditional uptake measurements.
ISSN:0033-7587
1938-5404
DOI:10.2307/3579012