Ionization in Solid- and Grid-Walled Detectors

Energy deposition spectra produced in a grid-walled proportional counter are compared to spectra produced in a conventional solid-walled detector. The radiation sources studied include 60 Co gamma rays, filtered 250 kVp X rays and unfiltered 250 kVp and 65 kVp X rays. The solid wall is shown to prod...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiat. Res. 51: No. 3, 569-80(Sep 1972) 569-80(Sep 1972), 1972-09, Vol.51 (3), p.569-580
Hauptverfasser: Braby, L. A., Ellett, W. H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Energy deposition spectra produced in a grid-walled proportional counter are compared to spectra produced in a conventional solid-walled detector. The radiation sources studied include 60 Co gamma rays, filtered 250 kVp X rays and unfiltered 250 kVp and 65 kVp X rays. The solid wall is shown to produce significant distortion of the spectra for 60 Co gamma rays and filtered 250 kVp X rays but not for the unfiltered X rays. The size of the container housing the grid-walled detector was varied and the wall effect was found to decrease less rapidly with increasing distance between the detector and the wall than would be expected on the basis of geometric considerations.
ISSN:0033-7587
1938-5404