A Stress Analysis of the Proximal Tibia after Insertion of the Endprosthesis
Detailed investigation for the local stress increment at the fixation post bottom of an artificial knee joint was performed by two-dimensional finite element method. A strain measurement was also conducted by wire strain gauges on a cancellous bone with a special method. It was made clear that a cur...
Gespeichert in:
Veröffentlicht in: | Keio journal of medicine 1989, Vol.38(3), pp.294-310 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Detailed investigation for the local stress increment at the fixation post bottom of an artificial knee joint was performed by two-dimensional finite element method. A strain measurement was also conducted by wire strain gauges on a cancellous bone with a special method. It was made clear that a curvature with a certain extent of radius was necessary for the corner of a fixation post and application of acrylic cement with a certain extent of thickness (3 mm≥) was effective for stress relaxation, and that an optimum value (10-110 GPa) existed for elastic modulus of bone cement. |
---|---|
ISSN: | 0022-9717 1880-1293 |
DOI: | 10.2302/kjm.38.294 |