Amorphous Silicon Thermometer

The carbon glass resistance thermometers (CGRT) shows an unstable drift by heat cycles. Since we were looking for a more stable element of thermometer for cryogenic and high magnetic field environments, we selected amorphous silicon as a substitute for CGRT. The resistance of many amorphous samples...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:TEION KOGAKU (Journal of Cryogenics and Superconductivity Society of Japan) 1988/06/25, Vol.23(3), pp.140-144
Hauptverfasser: YOSHIDA, Kiyoshi, YAMAURA, Yoshiro, TAWADA, Yoshihisa
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The carbon glass resistance thermometers (CGRT) shows an unstable drift by heat cycles. Since we were looking for a more stable element of thermometer for cryogenic and high magnetic field environments, we selected amorphous silicon as a substitute for CGRT. The resistance of many amorphous samples were measured at 4K, at 77K, and at 300K. We eventually found an amorphous silicon (Si-H) alloy whose the sensitivity below 77K was comparable to that of the germanium resistance thermometer with little magnetic field influence.
ISSN:0389-2441
1880-0408
DOI:10.2221/jcsj.23.140