Perclose ProGlide embolization as a complication: case report and review of literature

Vascular closure devices have replaced mechanical compression for closure of femoral access sites after endovascular procedures. We present an 87-year-old male with symptomatic infrarenal aortic aneurysm measuring 4.8 cm presenting for elective endovascular repair of the aortic aneurysm. A Perclose...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Future cardiology 2021-10, Vol.17 (7), p.1193-1197
Hauptverfasser: Giniyani, Larab L, Rana, Yesha P, J Hanumanthu, Balaram Krishna, Chan, Doris, Kwan, Tak W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Vascular closure devices have replaced mechanical compression for closure of femoral access sites after endovascular procedures. We present an 87-year-old male with symptomatic infrarenal aortic aneurysm measuring 4.8 cm presenting for elective endovascular repair of the aortic aneurysm. A Perclose ProGlide Suture-Mediated Closure was used for closure. The closure was complicated by a separation of the ProGlide device resulting in the migration of the footplate to the descending aorta. Correction required snare retrieval via radial access, and the patient recovered without complications. We highlight an important complication of the Perclose ProGlide Suture-Mediated Closure device that is rare but important to know when performing endovascular closures with this device.
ISSN:1479-6678
1744-8298
DOI:10.2217/fca-2020-0154