A proposal to minimize the dispersion on primary calibration results of single-ended accelerometers at high frequencies

Primary calibrations of single-ended accelerometers by laser interferometric methods can be strongly affected at high frequencies by the vibration exciter used to generate the mechanical motion. A study was carried out at the Vibration Laboratory of INMETRO aiming to better understand the causes of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Acta IMEKO 2014-01, Vol.2 (2), p.48
Hauptverfasser: Ripper, Gustavo Palmeira, Micheli, Giancarlo Barbosa, Dias, Ronaldo Da Silva
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Primary calibrations of single-ended accelerometers by laser interferometric methods can be strongly affected at high frequencies by the vibration exciter used to generate the mechanical motion. A study was carried out at the Vibration Laboratory of INMETRO aiming to better understand the causes of this dependence and to find a way to reduce its effect. A simple solution is proposed by the authors to reduce the dispersion on the sensitivity results evidenced when different vibration exciters are used. This proposal can be very helpful for calibration laboratories willing to compare results obtained in different measurement systems associated with low measurement uncertainties. Therefore, it might benefit processes, which includes for example the validation of calibration systems and interlaboratory comparisons.