9.2: Temporal and Color Measurements in Medical Displays

We report on the accurate characterization of temporal response, maximum color difference and luminance uniformity in medical liquid crystal displays. We find that the variability among results from different methods can be larger than measured color differences, and that poor response time estimate...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:SID International Symposium Digest of technical papers 2006-06, Vol.37 (1), p.97-100
Hauptverfasser: Liang, Hongye, Saha, Anindita, Badano, Aldo
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report on the accurate characterization of temporal response, maximum color difference and luminance uniformity in medical liquid crystal displays. We find that the variability among results from different methods can be larger than measured color differences, and that poor response time estimates occur for transitions between neighboring gray levels.
ISSN:0097-966X
2168-0159
DOI:10.1889/1.2433689