This paper describes the evaluation of a new method of sample preparation using a cold finger system with continuous water flow for rice analysis by flame atomic absorption spectrometry. The limits of detection for Cu, Fe, Mn and Zn for the proposed method were 0.36, 1.84, 2.12 and 0.16 mg kg-1, res...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Química Nova 2014, Vol.37 (1), p.6-9
Hauptverfasser: Pinheiro, Aline Colvara de Almeida, Lisboa, Meibel Teixeira, Ribeiro, Anderson Schwingel, Nunes, Adriane Medeiros, Yamasaki, Alzira
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper describes the evaluation of a new method of sample preparation using a cold finger system with continuous water flow for rice analysis by flame atomic absorption spectrometry. The limits of detection for Cu, Fe, Mn and Zn for the proposed method were 0.36, 1.84, 2.12 and 0.16 mg kg-1, respectively. The RSDs were lower than 6.0% for all elements and the CRM analyzed showed values with 95% agreement. The proposed method is simple and safe for the proposed objective and does not require the use of mixtures of acid or special equipment for sample preparation.