Simultaneous and accurate determination of water- and fat-soluble vitamins in multivitamin tablets by using an RP-HPLC method
In the present study, a reversed-phase high-performance liquid chromatographic (RP-HPLC) procedure was developed and validated for the simultaneous determination of seven water-soluble vitamins (thiamine, riboflavin, niacin, cyanocobalamin, ascorbic acid, folic acid, and p-aminobenzoic acid) and fou...
Gespeichert in:
Veröffentlicht in: | Química Nova 2013, Vol.36 (7), p.1044-1051 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In the present study, a reversed-phase high-performance liquid chromatographic (RP-HPLC) procedure was developed and validated for the simultaneous determination of seven water-soluble vitamins (thiamine, riboflavin, niacin, cyanocobalamin, ascorbic acid, folic acid, and p-aminobenzoic acid) and four fat-soluble vitamins (retinol acetate, cholecalciferol, α-tocopherol, and phytonadione) in multivitamin tablets. The linearity of the method was excellent (R² > 0.999) over the concentration range of 10 - 500 ng mL-1. The statistical evaluation of the method was carried out by performing the intra- and inter-day precision. The accuracy of the method was tested by measuring the average recovery; values ranged between 87.4% and 98.5% and were acceptable quantitative results that corresponded with the label claims. |
---|---|