A Calibrated Large-area X-Ray Source for Plasma Spectroscopy

A large-area (10 x 30 cm ) X -ray source for relative and absolute calibration of double-crystal monochromators for plasma soft X -ray spectroscopy was developed. For a voltage of 20 kV and a current of 1 mA uniform and reproducible emission is achieved at a level of about 30 mW /(m · sr) for K , li...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Zeitschrift für Naturforschung. A, A journal of physical sciences A journal of physical sciences, 1987-10, Vol.42 (10), p.1051-1053
Hauptverfasser: Morsi, H. W., Röhr, H., Schumacher, U.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A large-area (10 x 30 cm ) X -ray source for relative and absolute calibration of double-crystal monochromators for plasma soft X -ray spectroscopy was developed. For a voltage of 20 kV and a current of 1 mA uniform and reproducible emission is achieved at a level of about 30 mW /(m · sr) for K , line emission in the photon energy range from 1 keV to about 8 keV, while the L line emission increases with photon energy from 10 to 20 m W/(m · sr).