A New Stability Indicating Reverse Phase HPLC Method for the Determination of Related Substances of DPP4 Inhibitor Drug: Linagliptin
A simple, sensitive, selective and reproducible stability indicating reverse phase HPLC method has been developed and validated for the determination of related substances of linagliptin drug substance. Separation of impurities was carried out on a YMC Pack ODS-AQ, 5 μm (250 × 4.6 mm) column. Column...
Gespeichert in:
Veröffentlicht in: | Asian journal of chemistry 2022, Vol.34 (10), p.2587-2592 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A simple, sensitive, selective and reproducible stability indicating reverse phase HPLC method has
been developed and validated for the determination of related substances of linagliptin drug substance.
Separation of impurities was carried out on a YMC Pack ODS-AQ, 5 μm (250 × 4.6 mm) column.
Column eluent consists of gradient mixing of phosphate buffer at pH-3.7 as mobile phase A and a
mixture of acetonitrile, methanol and water in the ratio of 600:250:150 v/v/v as mobile phase-B.
Gradient elution at a flow rate of 1.0 mL/min and injection volume is 10 μL. The UV/vis detector is set
to a wavelength of 226 nm and the column oven is set at 30 ºC. The detector response was found to be
linear over the concentration range of 0.07-1.1 μg/mL with correlation coefficients greater than 0.999
for linagliptin and its related substances. LOD values for specified impurities range from 0.004-0.005%
w/w and LOQ values from 0.013-0.015% w/w. The accuracy obtained by the proposed method was
found to be in the range of 93.2-101.3%. Degraded and process impurities are well separated and the
method has been validated for specificity, precision, ruggedness, linearity, accuracy and robustness as
per ICH guidelines. |
---|---|
ISSN: | 0970-7077 0975-427X |
DOI: | 10.14233/ajchem.2022.23924 |