Dry Powder Inhalers Are Bioequivalent to Metered-Dose Inhalers

Metered-dose inhalers (MDIs) are extensively used to deliver drugs to the lungs but are driven by chlorofluorocarbon (CFC) propellants. The worldwide phasing out of CFCs within the next 5 to 10 years presents difficulties to the pharmaceutical industry. The mean ± SD relative lung bioavailability of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chest 1995-03, Vol.107 (3), p.629-633
Hauptverfasser: Hindle, Michael, Newton, Duncan A.G., Chrystyn, Henry
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Metered-dose inhalers (MDIs) are extensively used to deliver drugs to the lungs but are driven by chlorofluorocarbon (CFC) propellants. The worldwide phasing out of CFCs within the next 5 to 10 years presents difficulties to the pharmaceutical industry. The mean ± SD relative lung bioavailability of albuterol to the lung following inhalation of 400 µg of albuterol from an MDI, the Rotahaler and Diskhaler in 10 well-trained volunteers, was 2.83 (0.78), 1.72 (0.99), and 2.64 (1.23)%, respectively, expressed as a percentage of the nominal dose. The delivery of albuterol to the lungs from the MDI and Diskhaler was similar. In nine asthmatic subjects, the relative lung bioavailability of albuterol following inhalation with the MDI and Diskhaler was 1.19 (0.79) and 2.38 (1.46)%, respectively, expressed as a percentage of the nominal dose. There was no difference in reversibility 30 min after administration of the dose by the two methods. Similar lung deposition from the Diskhaler in volunteers probably is due to efficient MDI technique, which was absent in the asthmatic subjects. The Diskhaler does not rely on coordination during inhalation and therefore is easier to use.
ISSN:0012-3692
1931-3543
DOI:10.1378/chest.107.3.629