Low dose protocol for high resolution CT thorax: influence of matrix size and tube voltage on image quality and radiation dose [version 1; peer review: 1 approved, 1 approved with reservations]
Background: High-resolution CT (HRCT) thorax has increase demand due to its advantage in diagnosing chronic respiratory diseases. The feasibility of matrix size with different tube voltage in the HRCT protocol of thorax is unknown. Therefore, this study aimed to compare the effect of matrix sizes an...
Gespeichert in:
Veröffentlicht in: | F1000 research 2022, Vol.11, p.399 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background: High-resolution CT (HRCT) thorax has increase demand due to its advantage in diagnosing chronic respiratory diseases. The feasibility of matrix size with different tube voltage in the HRCT protocol of thorax is unknown. Therefore, this study aimed to compare the effect of matrix sizes and tube voltage on image quality and radiation dose on adult HRCT thorax.
Methods: A Phantom experiment was performed, followed by a patient scan. For phantom and patient scan, a total of six protocols with two tube voltage settings, 120 kVp and 100 kVp, with a combination of three matrix sizes, 512, 768, and 1024 were used. In this study, 180 adult patients who had HRCT thorax scan were considered. Dose data was collected, and quantitative image analysis was performed by drawing region of interests on the acquired phantom and patient images. Qualitative image analysis was performed independently by two blinded radiologists.
Results: The dose report of the phantom experiment revealed that the 100kVp with selected matrix size delivered 15.64% and 15.62% less radiation dose in terms of volumetric computed tomography dose index (CTDIvol) and dose length product (DLP), respectively, compared to 120kVp settings with selected matrix sizes. Similarly, for the patient population, the CTDIvol and DLP difference noted for 120kVp and 100kVp with different matrix sizes was statistically significant (p |
---|---|