Calibration frequency of dose--area product meters

Calibration of patient dose monitoring devices in diagnostic radiology has become increasingly important in the light of new legislation that requires monitoring of patient dose against local and national diagnostic reference levels. An investigation was conducted into the long-term stability of 41...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of radiology 2001-03, Vol.74 (879), p.259-261
Hauptverfasser: Crawley, M T, Mutch, S, Nyekiova, M, Reddy, C, Weatherburn, H
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Calibration of patient dose monitoring devices in diagnostic radiology has become increasingly important in the light of new legislation that requires monitoring of patient dose against local and national diagnostic reference levels. An investigation was conducted into the long-term stability of 41 dose-area product (DAP) meters over a period of approximately 5 years, to assess the suitability of an annual calibration regimen. For DAP meters fitted to overcouch X-ray tubes, 77% of calibrations were within 10%, whilst for undercouch tubes only 50% of calibrations were within 10%. These findings suggest that annual calibration may be too infrequent. Suitable calibration frequencies for different clinical workloads are discussed.