Investigation of optimum X-ray beam tube voltage and filtration for chest radiography with a computed radiography system

The purpose of this study was to determine the optimum tube voltage and amount of added copper (Cu) filtration for processed chest radiographs obtained with an Agfa 75.0 Computed Radiography (CR) system. The contrast-to-noise ratio (CNR) was measured in the lung, heart/spine and diaphragm compartmen...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of radiology 2008-10, Vol.81 (970), p.771-777
Hauptverfasser: MOORE, C. S, BEAVIS, A. W, SAUNDERSON, J. R
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to determine the optimum tube voltage and amount of added copper (Cu) filtration for processed chest radiographs obtained with an Agfa 75.0 Computed Radiography (CR) system. The contrast-to-noise ratio (CNR) was measured in the lung, heart/spine and diaphragm compartments of a validated chest phantom using various tube voltages and amounts of Cu filtration. The CNR was derived as a function of air kerma at the CR plate and with the effective dose. As rib contrast can interfere with detection of nodules in chest radiography, a tissue-to-rib ratio (TRR) was derived to investigate which tube voltages suppress the contrast of rib. Although processing algorithms affect the signal and noise in a way that is hard to predict, we found that, for a given set of processing parameters, the CNR was related to the plate air kerma and effective dose in a logarithmic manner (all R(2) >or=0.97). For imaging of the lung region, a low voltage (60 kVp) produced the highest CNR, whereas a high voltage (125 kVp) produced the highest TRR. In the heart/spine region, 80-125 kVp produced the highest CNR, while in the diaphragm region 60-90 kVp produced the highest CNR. For chest radiography with this CR system, the optimal tube voltage depends upon the region of interest. Of the filters tested, a 0.1 mm Cu thickness was found to provide a statistically significant increase in the CNR in the diaphragm region with tube potentials of 60 kVp and 80 kVp, without affecting the CNR in the other anatomical compartments.
ISSN:0007-1285
1748-880X
DOI:10.1259/bjr/21963665