Improvements in the Technique of Spinal Subarachnoid Recirculatory Perfusion

With /sup 14/C-labeled dextran as the tracer, studies of the original configuration of spinal recirculatory perfusion and the original model for data analysis demonstrated that this approach does not yield acceptable accuracy in determining cerebrospinal fluid (CSF) formation (Fcsf) and absorption (...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Neurosurgery 1984-08, Vol.15 (2), p.203-213
Hauptverfasser: Sullivan, Humbert G., Allison, Jerry D., Kingsbury, Theodore B., Goode, Jamie J., Sims, William L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:With /sup 14/C-labeled dextran as the tracer, studies of the original configuration of spinal recirculatory perfusion and the original model for data analysis demonstrated that this approach does not yield acceptable accuracy in determining cerebrospinal fluid (CSF) formation (Fcsf) and absorption (Acsf) rates. A significant component of this error was due to the fact that the method of data analysis used originally is not based on a realistic mathematical model of the system. A more realistic mathematical model resulted in two simultaneous differential equations that did not have simple analytical solutions and, therefore, could not be used easily for data analysis. By computer simulation, a comparison of the more realistic model with the original model demonstrated that, under ideal conditions, there was a 10% error inherent in the original data analysis method. A new system configuration and new data analysis methods have been developed. In the new system, the syringe is removed from the external circuit and intracranial pressure is controlled by infusion from a separate reservoir where the pressure head is maintained at any desired level by feedback control. Spectrophotometry is used to measure tracer concentration in the external circuit. Data collection and analysis are fully automated under computer control so that, during an experimental run, the investigators are updated at 1- to 2-second intervals as to the convergence of the data analysis routine. All of the data during the initial period of nonhomogeneous mixing are used to calculate Fcsf and Acsf. With the new system and a simple phantom of the CSF system, the mean error in finding Acsf was 1.7 +/- 1.2% for 27 determinations covering a wide range of absorption rates. Fcsf could be determined to within 0.001 ml/minute. In up to six sequential pressure plateaus, the magnitude of error did not increase with each subsequent run.
ISSN:0148-396X
DOI:10.1227/00006123-198408000-00010