Peripheral bronchial luminal conspicuity on dynamic-ventilation computed tomography: association with radiation doses and temporal resolution by using an ex vivo porcine lung phantom
Background It is still unclear which image reconstruction algorithm is appropriate for peripheral bronchial luminal conspicuity (PBLC) on dynamic-ventilation computed tomography (DVCT). Purpose To assess the influence of radiation doses and temporal resolution (TR) on the association between movemen...
Gespeichert in:
Veröffentlicht in: | Acta radiologica (1987) 2020-12, Vol.61 (12), p.1608-1617 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
It is still unclear which image reconstruction algorithm is appropriate for peripheral bronchial luminal conspicuity (PBLC) on dynamic-ventilation computed tomography (DVCT).
Purpose
To assess the influence of radiation doses and temporal resolution (TR) on the association between movement velocity (MV) and PBLC on DVCT.
Material and Methods
An ex vivo porcine lung phantom with simulated respiratory movement was scanned by 320-row CT at 240 mA and 10 mA. Peak and dip CT density and luminal area adjusted by values at end-inspiration (CTDpeak and CTDdip, luminal area ratio [LAR]) for PBLC and MVs were measured and visual scores (VS) were obtained at 12 measurement points on 13 frame images obtained at half and full reconstructions (TR 340 and 190 ms) during expiration. Size-specific dose estimate (SSDE) was applied to presume radiation dose. VS, CTDpeak, CTDdip, LAR, and their cross-correlation coefficients with MV (CCC) were compared among four methods with combinations of two reconstruction algorithms and two doses.
Results
The dose at 10 mA was presumed as 26 mA by SSDE for standard proportion adults. VS, CTDdip, CTDpeak, and LAR with half reconstruction at 10 mA (2.52 ± 0.59, 1.016 ± 0.221, 0.948 ± 0.103, and 0.990 ± 0.527) were similar to those at 240 mA except for VS, and different from those with full reconstruction at both doses (2.24 ± 0.85, 0.830 ± 0.209, 0.986 ± 0.065, and 1.012 ± 0.438 at 240 mA) (P |
---|---|
ISSN: | 0284-1851 1600-0455 |
DOI: | 10.1177/0284185120911186 |