Optoacoustic Detection of Microparticles in Liquids at LaserFluences Below the Optical Breakdown Threshold

An optoacoustic device consisting of a XeCl excimer laser and a measurement cell with an attached piezotransducer was used for detecting microparticles suspended in liquid probes. The potential of optoacoustic diagnostics of microinhomogeneous liquids was enhanced by applying informative parameters...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Laser chemistry 1995-01, Vol.16 (2), p.63-73
Hauptverfasser: Alimpiev, S. S., Simanovskii, Ya. O., Egerev, S. V., Pashin, A. E.
Format: Artikel
Sprache:eng ; jpn
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An optoacoustic device consisting of a XeCl excimer laser and a measurement cell with an attached piezotransducer was used for detecting microparticles suspended in liquid probes. The potential of optoacoustic diagnostics of microinhomogeneous liquids was enhanced by applying informative parameters of optoacoustic response. Probes of distilled water and Dow Chemical latex suspension were used in experiments. It was shown that cavitation mechanism of optoacoustic conversion taking place at laser fluences below the optical breakdown threshold was well‐suited for individual microparticle detection. The approach proposed is different from the well‐known breakdown counting optoacoustic technique.
ISSN:0278-6273
1476-3516
DOI:10.1155/1995/43684