RF Attenuation Characteristics for In Vivo Wireless Healthcare Chip
We investigate a wireless communication system for an in vivo healthcare chip. In this paper, we present measured attenuation characteristics through the human body at several frequencies. In the measurement, we use physiological saline and fresh meat instead of a real human body. From the measured...
Gespeichert in:
Veröffentlicht in: | Japanese Journal of Applied Physics 2005-07, Vol.44 (7R), p.5275 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We investigate a wireless communication system for an
in vivo
healthcare chip. In this paper, we present measured attenuation characteristics through the human body at several frequencies. In the measurement, we use physiological saline and fresh meat instead of a real human body. From the measured results, we found that 13.56 MHz has an attenuation of 47 dB and is suitable for the proposed system. |
---|---|