Impurity-doped GaSe radiation detector evaluated at 100°C
The pulse height distribution for 5.5 MeV α particles from 241 Am was measured using a GaSe detector doped with 0.1 at.% Sn at 100°C and compared with that measured at room temperature. It was found that the energy resolution at 100°C was 8.3% which was as good as that at room temperature. The tempe...
Gespeichert in:
Veröffentlicht in: | Japanese Journal of Applied Physics 1997, Vol.36 (1A), p.378-379 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The pulse height distribution for 5.5 MeV α particles from
241
Am was measured using a GaSe detector doped with 0.1 at.% Sn at 100°C and compared with that measured at room temperature. It was found that the energy resolution at 100°C was 8.3% which was as good as that at room temperature. The temperature dependence of the leakage current of the detector was also measured for the applied voltage of 50 V and found to have a small value of 60 nA at 100°C. |
---|---|