Estimation of Initial and Response Times of Laser Dew-Point Hygrometer by Measurement Simulation
The initial and the response times of the laser dew-point hygrometer were evaluated by measurement simulation. The simulation was based on loop computations of the surface temperature of a plate with dew deposition, the quantity of dew deposited and the intensity of scattered light from the surface...
Gespeichert in:
Veröffentlicht in: | Japanese Journal of Applied Physics 1995, Vol.34 (10), p.5847-5852, Article 5847 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The initial and the response times of the laser dew-point hygrometer were evaluated by measurement simulation. The simulation was based on loop computations of the surface temperature of a plate with dew deposition, the quantity of dew deposited and the intensity of scattered light from the surface at each short interval of measurement. The initial time was defined as the time necessary for the hygrometer to reach a temperature within ±0.5° C of the measured dew point from the start time of measurement, and the response time was also defined for stepwise dew-point changes of +5° C and -5° C. The simulation results are in approximate agreement with the recorded temperature and intensity of scattered light of the hygrometer. The evaluated initial time ranged from 0.3 min to 5 min in the temperature range from 0° C to 60° C, and the response time was also evaluated to be from 0.2 min to 3 min. |
---|---|
ISSN: | 0021-4922 1347-4065 |
DOI: | 10.1143/JJAP.34.5847 |