Stability-indicating high performance thin layer chromatographic determination of sulfanilamide in human urine
A simple, sensitive, selective, precise and stability-indicating high-performance thin-layer chromatographic ( HPTLC ) method was developed and applied to human urine for the densitometric determination of sulfanilamide. A mixture of chloroform-ethyl acetate-xylene (2.5: 4.0: 1.0, v/v/v) was used as...
Gespeichert in:
Veröffentlicht in: | Journal of analytical chemistry (New York, N.Y.) N.Y.), 2009-07, Vol.64 (7), p.714-720 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A simple, sensitive, selective, precise and stability-indicating high-performance thin-layer chromatographic (
HPTLC
) method was developed and applied to human urine for the densitometric determination of sulfanilamide. A mixture of chloroform-ethyl acetate-xylene (2.5: 4.0: 1.0, v/v/v) was used as a mobile phase. The system was found to give compact spots for sulfanilamide (retardation factor,
R
f
= 0.21±0.02). The linear regression analysis data for the calibration plots showed good linear relationship with
r
2
= 0.9970 ± 0.0003 and
r
2
= 0.9947 ± 0.020 within the concentration range of 50–250 ng per spot and 100–1000 ng per spot with respect to peak area, respectively. The limit of detection (
LOD
) and quantification (
LOQ
) were 8 and 25 ng per spot, respectively. Sulfanilamide was subjected to acid and alkali hydrolysis, oxidation, dry heat and wet heat treatment. According to the International Conference on Harmonization (
ICH
) guidelines the method was validated for precision, recovery and robustness. The ultraviolet (
UV
) spectra of the degradation products which had different spectra from sulfanilamide were also recorded. |
---|---|
ISSN: | 1061-9348 1608-3199 |
DOI: | 10.1134/S1061934809070090 |