An instrument for measuring the intensity and dose of cancerigenic ultraviolet radiation

An instrument for measuring the intensity and dose of cancerigenic radiation from the Sun and artificial sources has been developed. A 4H-SiC Schottky barrier photodetector is used as the sensor. The sensitivity spectrum of the photoelectric transducer lies in the region 240–300 nm and has a maximum...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Instruments and experimental techniques (New York) 2010-09, Vol.53 (5), p.743-745
Hauptverfasser: Blank, T. V., Gol’dberg, Yu. A., Kalinina, E. V., Karpenko, A. N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An instrument for measuring the intensity and dose of cancerigenic radiation from the Sun and artificial sources has been developed. A 4H-SiC Schottky barrier photodetector is used as the sensor. The sensitivity spectrum of the photoelectric transducer lies in the region 240–300 nm and has a maximum at 260 nm. The quantum efficiency is ≈0.3 electrons/photon (at λ = 254 nm), and its temperature coefficient at T = 250–310 K is < 0.1%/°C. The dimensions of the device are 3 × 6 × 11 cm.