Extrapolated surface dose measurements with radiochromic film

A radiochromic film extrapolation method is described for the measurement of surface dose from high energy photon beams. Extrapolated central axis entrance surface dose using Gafchromic film for a 10×10  cm 2 field size is 15%±2% and 13%±2% of D max for 6 and 10 MV x rays, respectively. Extrapolated...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 1999-03, Vol.26 (3), p.485-488
Hauptverfasser: Butson, Martin J., Yu, Peter K. N., Metcalfe, Peter E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A radiochromic film extrapolation method is described for the measurement of surface dose from high energy photon beams. Extrapolated central axis entrance surface dose using Gafchromic film for a 10×10  cm 2 field size is 15%±2% and 13%±2% of D max for 6 and 10 MV x rays, respectively. Extrapolated surface dose for a 30×30  cm 2 field with a 10 mm perspex block tray is 49%±2% and 48%±2% of D max for 6 and 10 MV beams, respectively. All results agree with uncorrected Attix parallel plate ionization chamber surface ionization within 4% for the same beam energies and configurations.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.598539