Application of a radiophotoluminescent glass plate dosimeter for small field dosimetry
We have recently developed a prototypical radiophotoluminescent glass plate dosimeter (GPD) system as a device for small field dosimetry. The purpose of this study is to examine the usefulness of the GPD system for small field dosimetry. The profiles measured with the GPD were evaluated by comparing...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2005-06, Vol.32 (6), p.1548-1554 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We have recently developed a prototypical radiophotoluminescent glass plate dosimeter (GPD) system as a device for small field dosimetry. The purpose of this study is to examine the usefulness of the GPD system for small field dosimetry. The profiles measured with the GPD were evaluated by comparing them to those from Kodak X-Omat V and GAFCROMIC XR type R film dosimeters for 2, 5, 9, and
15
mm
circular collimators created by a linear accelerator-based radiosurgery system. The GPD output factors were compared with those of various detectors including an ion chamber, a
p
-type silicon diode detector, a glass rod dosimeter (GRD), and a diamond detector. The results measured with the GPD were also confirmed by comparing them to those from Monte Carlo simulations. The accuracy of a simulated beam is validated by the excellent agreement between Monte Carlo calculated and measured central axis depth-dose curves for 9- and
15
mm
circular collimators using 4- and
10
MV
photon beams. The GPD profiles show almost the same full width at half maximum as those of film dosimeters and Monte Carlo simulations at 4- and
10
MV
photon beams, but a little narrower penumbrae than the film dosimeters and Monte Carlo simulations. The output factors measured with the GPD are in good agreement with those from a diode detector, a diamond detector, and the GRD with a small active volume and Monte Carlo simulations, except for a very small
2
mm
circular collimator. It was found that the GPD is a very useful detector for small field dosimetry. |
---|---|