Technical Note: Output and energy fluctuations of the tomotherapy Hi-Art helical tomotherapy system
The output and energy calibrations for the first clinical Hi-Art 2.0 helical tomotherapy system have been reviewed. Fixed-gantry/fixed-couch and rotational-gantry/fixed-couch measurements were made on a daily basis over a period of 20 weeks to investigate system stability. Static gantry measurements...
Gespeichert in:
Veröffentlicht in: | Medical physics (Lancaster) 2004-07, Vol.31 (7), p.2119-2120 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The output and energy calibrations for the first clinical Hi-Art 2.0 helical tomotherapy system have been reviewed. Fixed-gantry/fixed-couch and rotational-gantry/fixed-couch measurements were made on a daily basis over a period of 20 weeks to investigate system stability. Static gantry measurements were taken at 10 cm depth in a rectangular stack of Virtual Water at an SSD distance of 90 cm and a field size of
5×40
cm
.
Rotational gantry measurements were taken in a cylindrical phantom Virtual Water phantom for a field size of
5×40
cm
.
The Hi-Art 2.0 system has maintained its calibration to within
±2%
and energy to within
±1.5%
over the initial 20 week period. |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1118/1.1763007 |