The x-ray sensitivity of amorphous selenium for mammography

A study of the x-ray sensitivity of amorphous selenium (a-Se) for digital mammography has been performed. A uniform layer of a-Se was deposited on a glass substrate with electrodes on both surfaces. The deposition procedure was identical to that used for a-Se flat-panel detectors. A high voltage was...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2002-03, Vol.29 (3), p.319-324
Hauptverfasser: Stone, Mary F., Zhao, Wei, Jacak, Barbara V., O’Connor, Paul, Yu, Bo, Rehak, Pavel
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A study of the x-ray sensitivity of amorphous selenium (a-Se) for digital mammography has been performed. A uniform layer of a-Se was deposited on a glass substrate with electrodes on both surfaces. The deposition procedure was identical to that used for a-Se flat-panel detectors. A high voltage was applied to the top surface of the a-Se layer in order to establish an electric field E Se . Then the sample was exposed to x rays with 27 kVp spectra generated from an x-ray tube with a molybdenum (Mo) target. The mean x-ray energy of the spectrum used was ∼16.6 keV. The x-ray current generated by the a-Se layer was measured as a function of E Se . From the current measurement and the estimation of total x-ray energy absorbed in the a-Se, the energy required to create one electron-hole pair (EHP), W, was determined as a function of E Se . It was found that at the most commonly used E Se of 10 V/μm, W was measured as 64 eV. This is considerably higher than the widely accepted typical value of W=50  eV measured at higher x-ray photon energies (e.g., 50 keV). The dependence of W as a function of E Se can be best fitted using the empirical expression of E Se −γ . This relationship is consistent with the results obtained at higher x-ray energies. This article provides an accurate measurement of x-ray sensitivity of a-Se at mammographic energies independent of detector operation, such as the most recently developed flat-panel detectors. The results will be a useful tool for investigation and optimization of a-Se-based x-ray imaging detectors, such as determination of pixel fill-factor and optimal E Se during operation.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.1449874