Axial-symmetric transmission gauge: Extension of its pressure measuring range and reduction of the electron stimulated desorption ion effect in ultrahigh vacuum

The basic characteristics of the axial-symmetric transmission gauge (AT gauge), that were modified for measurement over a wide pressure range, are reported. A Faraday cup ion collector is used instead of a secondary electron multiplier to extend the measurable pressure range to 10 − 3 Pa . The sensi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of vacuum science & technology. A, Vacuum, surfaces, and films Vacuum, surfaces, and films, 2005-05, Vol.23 (3), p.554-558
Hauptverfasser: Takahashi, N., Yuyama, J., Tuzi, Y., Akimichi, H., Arakawa, I.
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The basic characteristics of the axial-symmetric transmission gauge (AT gauge), that were modified for measurement over a wide pressure range, are reported. A Faraday cup ion collector is used instead of a secondary electron multiplier to extend the measurable pressure range to 10 − 3 Pa . The sensitivity of the gauge was determined by a conductance modulation method, and by direct comparison against an extractor gauge and a spinning rotor gauge as well. When operating in O 2 , there was systematic disagreement between pressure readings with the AT gauge and the extractor gauge. This is due to the difference in the elimination efficiency of the electron stimulated desorption ion.
ISSN:0734-2101
1520-8559
DOI:10.1116/1.1901668