A Low-Cost, Easily Deployable Vesicovaginal Fistula Occluding Device for Providing Interim Continence

Vesicovaginal fistulas (VVFs), abnormal openings between the vagina and bladder, disrupt the lives of over 3 million people worldwide due to resulting incontinence and infections. VVFs are commonly treated with surgery after the fistula has had time to heal over several months. In low-resource areas...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of medical devices 2022-06, Vol.16 (2)
Hauptverfasser: Das, Madhurima, Hahm, Katie S, LaRocca, Ava A, Luna, Cecilia Alessandra, Mendez, Keegan, Hoffman, Rachel, Verheyen, Connor A, Kim, Tesia G, Wamakima, Bridgette W, Roche, Ellen T
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Vesicovaginal fistulas (VVFs), abnormal openings between the vagina and bladder, disrupt the lives of over 3 million people worldwide due to resulting incontinence and infections. VVFs are commonly treated with surgery after the fistula has had time to heal over several months. In low-resource areas, immediate incontinence often leads to ostracization from the community and can be devastating for the patient. To address this problem, we have designed a three-tiered silicone plug consisting of a bladder-dwelling disk, a midfistula disk, and a vagina-dwelling cross-shaped tapered plug, all supported on a central stem. The system is designed to occlude the fistula and enable full continence until the patient is able to access surgery. This proof-of-concept device withstands typical expulsion forces from the bladder and does not leak under typical bladder filling or urination pressures. The maximum device expulsion force is 3.69 N and it is watertight up to 9.8 kPa. It is designed to be easily deployed by trained community members.
ISSN:1932-6181
1932-619X
DOI:10.1115/1.4053603