The Totaltrack TM video laryngeal mask: an evaluation in 300 patients
The Totaltrack is a new video laryngeal mask, which combines a supraglottic airway with a videolaryngoscope. We evaluated the performance of this device in an observational study in 300 adult patients. The Totaltrack was inserted at the first attempt and the glottis seen in all cases. In 249 patient...
Gespeichert in:
Veröffentlicht in: | Anaesthesia 2019-06, Vol.74 (6), p.751-757 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Totaltrack
is a new video laryngeal mask, which combines a supraglottic airway with a videolaryngoscope. We evaluated the performance of this device in an observational study in 300 adult patients. The Totaltrack was inserted at the first attempt and the glottis seen in all cases. In 249 patients (83%; 95%CI 78.7-87.3%) the glottis was seen at the first attempt. Adequate ventilation and tracheal intubation were achieved in all patients. The median time to visualisation of vocal cords, to confirmation of ventilation and to successful tracheal intubation was 5 s, 13 s and 24 s, respectively. Nineteen patients (6.3%; 95%CI 3.55-9.05%) suffered minor complications, such as mucosal lesion or blood staining. The Totaltrack appears effective for airway management. |
---|---|
ISSN: | 0003-2409 1365-2044 |
DOI: | 10.1111/anae.14637 |