Amplifier with time-invariant trapezoidal shaping and shape-sensitive pileup rejector for high-rate spectroscopy
An amplifier with trapezoidal pulse shaping was developed for high-rate high-energy gamma spectroscopy using NaI(Tl) scintillation detectors. It used a double-delay-line technique for producing a nearly triangular pulse shape combined with a linear circuit for producing a flat-topped pulse. Good ene...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on nuclear science 1989-02, Vol.36 (1), p.1326-1329 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An amplifier with trapezoidal pulse shaping was developed for high-rate high-energy gamma spectroscopy using NaI(Tl) scintillation detectors. It used a double-delay-line technique for producing a nearly triangular pulse shape combined with a linear circuit for producing a flat-topped pulse. Good energy resolution and short resolution time make this amplifier especially suitable for high-count-rate gamma-ray spectroscopy. To provide a versatile high-performance system, it includes a pileup rejector based on inspection of a pileup signal obtained by combining the slow output signal and fast-shaped input signal. The trapezoidal shape provides a short resolution time for minimal occurrence of pileup with a width suitable for presentation to a standard multichannel analyzer. The performance of the system was tested, and the results are presented.< > |
---|---|
ISSN: | 0018-9499 1558-1578 |
DOI: | 10.1109/TNS.1989.574132 |