A New High Sensitivity Thyroid X-Ray Fluorescence Scanning System

An array of five Si(Li) detectors which partially surrounds a collimated 10 Ci 241Am source, combined with a microprocessor-controlled scanning bed, has been developed for imaging of stable iodine within the thyroid. The sensitivity of the system is sufficient to generate high-quality scan images in...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on nuclear science 1981-01, Vol.28 (1), p.184-188
Hauptverfasser: Yasillo, N. J., Ortega, C. J., Charleston, D. B., Beck, R. N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An array of five Si(Li) detectors which partially surrounds a collimated 10 Ci 241Am source, combined with a microprocessor-controlled scanning bed, has been developed for imaging of stable iodine within the thyroid. The sensitivity of the system is sufficient to generate high-quality scan images in a scanning time of approximately 15 minutes, with a spatial resolution of 5.9 mm (FWHM). We have evaluated several phantom studies and used the results to develop a computer code which determines the total thyroid iodine content and thyroid area with acceptable accuracy.