The technique presented is an approach to absorp tion edge transmission imaging which eliminates the need for monoenergetic x-ray sources, detector scanning, and energy analysis of transmitted x-rays. Subtraction of images formed with heavily filtered x-ray beams peaked at energies above and below t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on nuclear science 1974-01, Vol.21 (1), p.645-651
Hauptverfasser: Mistretta, C. A., Kelcz, F., Ort, M. G., Siedband, M. P., Cameron, J. R., Crummy, A. B., Polcyn, R. E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The technique presented is an approach to absorp tion edge transmission imaging which eliminates the need for monoenergetic x-ray sources, detector scanning, and energy analysis of transmitted x-rays. Subtraction of images formed with heavily filtered x-ray beams peaked at energies above and below the K absorption edge is accomplished with image-intensified fluoroscopy equipment coupled to a pair of storage tubes. Images of iodine and xenon concentrations of 1 mg/cm2 have been imaged in 15 cm phantoms in a few seconds. When automated, the system will provide similar sensitivity in 20 cm phantoms, with imaging times of 1 second or less with resolution on the order of 1 mm. Several preliminary images are presented which relate to possible future clinical studies.