Iodine stabilized dye laser system for frequency measurements in the visible and near IR region of the spectrum
An iodine stabilized dye laser system is described that provides traceable measurement of reference frequencies in the visible spectrum from 540 to 670 nm and in the near infrared at 1.15 /spl mu/m. The system allows calibration of the widely used 633 nm, 612 nm, and 543 nm HeNe laser systems. Also,...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on instrumentation and measurement 2003-04, Vol.52 (2), p.272-276 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An iodine stabilized dye laser system is described that provides traceable measurement of reference frequencies in the visible spectrum from 540 to 670 nm and in the near infrared at 1.15 /spl mu/m. The system allows calibration of the widely used 633 nm, 612 nm, and 543 nm HeNe laser systems. Also, frequency measurements of a polarization stabilized 1153 nm HeNe laser have been performed via frequency doubling and comparison with the dye system operating on the corresponding 576 nm lines. Studies of the shift sensitivities of the system at various wavelengths of interest are described for variation of iodine cell pressure, laser modulation amplitude, and optical saturation power. The dye system was also stabilized to hyperfine components associated with the 6-3 P(33) iodine transition and compared with a 633 nm iodine stabilized HeNe standard. |
---|---|
ISSN: | 0018-9456 1557-9662 |
DOI: | 10.1109/TIM.2003.811662 |