Developing AIMD Models Using Orthogonal Pathways for MRI Safety Assessment

In this article, a method for active implantable medical devices (AIMD) model development is presented. Based on the transmission line model, the AIMD model can be developed semianalytically using a few direct measurements inside the American Society for Testing and Materials (ASTM) phantom. Folded...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on electromagnetic compatibility 2020-12, Vol.62 (6), p.2689-2695
Hauptverfasser: Zheng, Jianfeng, Wang, Zhichao, Wang, Qingyan, Hu, Sheng, Gu, Zhengyu, Kainz, Wolfgang, Chen, Ji
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this article, a method for active implantable medical devices (AIMD) model development is presented. Based on the transmission line model, the AIMD model can be developed semianalytically using a few direct measurements inside the American Society for Testing and Materials (ASTM) phantom. Folded orthogonal pathways based on the Hadamard matrix are used to make the problem of the AIMD model development to be well-conditioned. Both induced voltage and heating models for the example AIMDs were developed to demonstrate the effectiveness of this method.