Noise Analysis and Performance Comparison of Low Current Measurement Systems for Biomedical Applications

In this paper, we report on the noise analysis of low current measurement systems for biomedical applications and their fundamental limits. We analyzed resistive feedback, capacitive feedback and current amplifier circuits for low current measurement systems. Detailed noise analysis for different bi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on biomedical circuits and systems 2013-02, Vol.7 (1), p.52-62
Hauptverfasser: Dongsoo Kim, Goldstein, B., Wei Tang, Sigworth, F. J., Culurciello, E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this paper, we report on the noise analysis of low current measurement systems for biomedical applications and their fundamental limits. We analyzed resistive feedback, capacitive feedback and current amplifier circuits for low current measurement systems. Detailed noise analysis for different biomedical applications are presented and matched with measurement data using a 0.5-μm fabrication process. Based on the theoretical analysis and the corresponding measurement results, the capacitive feedback system provides better noise performance for the measurement of low current than the others. The capacitive feedback circuit is capable of measuring 750 fA RMS at a 10 kHz sampling rate, whereas the resistive feedback provides 4 pA and the current conveyor provides 600 pA at the same bandwidth. This paper provides design guidelines to maximize the performance of low current measuring system for biomedical instrumentation and to provide the best performance available with CMOS technologies.
ISSN:1932-4545
1940-9990
DOI:10.1109/TBCAS.2012.2192273