Comparison between the magnetization obtained from Faraday balance and VSM or SQUID

The authors derive a relation between the magnetization obtained using a vibrating sample magnetometer (VSM) or a superconducting quantum interference device (SQUID) magnetometer and the magnetization obtained using a Faraday balance (FB). It is shown that, to a first approximation, the magnetizatio...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on magnetics 1992-09, Vol.28 (5), p.2453-2455
Hauptverfasser: Dababneh, M., Odeh, I., El-Ali, A., Ayoub, N.Y.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The authors derive a relation between the magnetization obtained using a vibrating sample magnetometer (VSM) or a superconducting quantum interference device (SQUID) magnetometer and the magnetization obtained using a Faraday balance (FB). It is shown that, to a first approximation, the magnetization obtained using a FB should be very close to those obtained using VSM or SQUID, especially for short samples and low field. The variation of the magnetization with the sample length is presented at low and intermediate fields.< >
ISSN:0018-9464
1941-0069
DOI:10.1109/20.179523