Shielded open-circuited sample holder for dielectric measurements of solids and liquids

This paper presents a technique for permittivity measurements that can be used for either liquids or solid coaxial samples from near dc to gigahertz frequencies. The method uses both a full-mode model of a coaxial open-circuited termination and a capacitive method for permittivity determination. The...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on instrumentation and measurement 1998-04, Vol.47 (2), p.338-344
Hauptverfasser: Baker-Jarvis, J., Janezic, M.D., Jones, C.A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper presents a technique for permittivity measurements that can be used for either liquids or solid coaxial samples from near dc to gigahertz frequencies. The method uses both a full-mode model of a coaxial open-circuited termination and a capacitive method for permittivity determination. The method requires the use of an impedance bridge and a network analyzer. Measurements indicate good accuracy for the permittivity over a very wide band of frequencies. For high permittivity materials the sample surfaces must be metallized. An uncertainty analysis is also presented.
ISSN:0018-9456
1557-9662