Comparison of Parasitic Scattering from Window Materials used for Small-Angle X-ray Scattering: a Better Beryllium Window
The parasitic scattering of several materials used as windows in small‐angle scattering experiments, including IF‐1 beryllium, Kapton, Mylar, mica and glass has been measured over the q range 0.05–2.0 nm−1 with Cu Kα radiation (q = 4πsinθ/λ, 2θ = scattering angle). Based on the criteria of high stre...
Gespeichert in:
Veröffentlicht in: | Journal of applied crystallography 1995-12, Vol.28 (6), p.820-826 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The parasitic scattering of several materials used as windows in small‐angle scattering experiments, including IF‐1 beryllium, Kapton, Mylar, mica and glass has been measured over the q range 0.05–2.0 nm−1 with Cu Kα radiation (q = 4πsinθ/λ, 2θ = scattering angle). Based on the criteria of high strength, high transmission and low parasitic scattering, the best material was found to be IF‐1 beryllium. |
---|---|
ISSN: | 1600-5767 0021-8898 1600-5767 |
DOI: | 10.1107/S0021889895005784 |