A method of correcting for linear drift in computed tomography brain scans
Linear drift of X-ray attenuation coefficients must be corrected if quantitative comparisons are to be made between computed tomography (CT) brain scans of the same individual performed at different times. Such a correction is accomplished by comparing the low (cerebrospinal fluid) end of the attenu...
Gespeichert in:
Veröffentlicht in: | Journal of computer assisted tomography 1979-04, Vol.3 (2), p.251-255 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Linear drift of X-ray attenuation coefficients must be corrected if quantitative comparisons are to be made between computed tomography (CT) brain scans of the same individual performed at different times. Such a correction is accomplished by comparing the low (cerebrospinal fluid) end of the attenuation coefficient frequency histograms using a percentile--percentile plot. A "drift correction" permits serial quantitative assessments of the progression or regression of white matter hypodensity, such as occurs in drug induced leukoencephalopathy. |
---|---|