A method of correcting for linear drift in computed tomography brain scans

Linear drift of X-ray attenuation coefficients must be corrected if quantitative comparisons are to be made between computed tomography (CT) brain scans of the same individual performed at different times. Such a correction is accomplished by comparing the low (cerebrospinal fluid) end of the attenu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of computer assisted tomography 1979-04, Vol.3 (2), p.251-255
Hauptverfasser: Thaler, H T, Rottenberg, D A, Pentlow, K S, Allen, J C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Linear drift of X-ray attenuation coefficients must be corrected if quantitative comparisons are to be made between computed tomography (CT) brain scans of the same individual performed at different times. Such a correction is accomplished by comparing the low (cerebrospinal fluid) end of the attenuation coefficient frequency histograms using a percentile--percentile plot. A "drift correction" permits serial quantitative assessments of the progression or regression of white matter hypodensity, such as occurs in drug induced leukoencephalopathy.
ISSN:0363-8715
DOI:10.1097/00004728-197904000-00019