Personal dosimetry in terms of HP(3):Monte Carlo and experimental studies
Hp(3) has been defined as the operational quantity for eye lens dosimetry. Hp(3)/ka conversion coefficients were evaluated at the GSF (Germany) in a 30×30×15 cm3 4-elements ICRU slab phantom for various energies and incident angles through Monte Carlo. The ISO report 12,794 suggests to employ a PMMA...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2007, Vol.125 (1-4), p.145-148 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Hp(3) has been defined as the operational quantity for eye lens dosimetry. Hp(3)/ka conversion coefficients were evaluated at the GSF (Germany) in a 30×30×15 cm3 4-elements ICRU slab phantom for various energies and incident angles through Monte Carlo. The ISO report 12,794 suggests to employ a PMMA water filled phantom, of the same dimensions, for dosemeter calibration in terms of Hp(3). The present paper briefly summarises the main aspects of a study carried out at ENEA-Radiation Protection Institute (Bologna, Italy) to provide practical procedures for the calibration of dosemeters in terms of Hp(3). Tabulations of a new set conversion coefficients and air kerma backscatter factors are provided as a function of energy and incident angle. The paper demonstrates that a more accurate approach to the dosimetric assessment in terms of Hp(3) could be rather simply introduced employing a reduced phantom. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncl385 |