Analysis of image quality in digital chest imaging
An evaluation of the image quality of an amorphous silicon flat-panel detector system and a computed radiology system compared with a screen-film system was performed by means of contrast-detail phantom images. Hard and soft copy images were evaluated. Although patient dose at clinical settings was...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2005-01, Vol.117 (1-3), p.174-177 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An evaluation of the image quality of an amorphous silicon flat-panel detector system and a computed radiology system compared with a screen-film system was performed by means of contrast-detail phantom images. Hard and soft copy images were evaluated. Although patient dose at clinical settings was strongly decreased with the amorphous silicon system, the low-contrast visibility with this system was still significantly better than with the screen-film system. For the computed radiology system, low-contrast visibility was comparable to the screen-film system. Best results were obtained by soft copy reading at full resolution with adaptation of contrast and brightness. Changing tube voltage (102–133 kV), or additional filtration, did not significantly affect image quality. However, low-contrast visibility improved significantly with increasing exposure. It was clearly demonstrated that, in chest imaging, the amorphous silicon system has superior imaging characteristics compared to the screen-film and the computed radiology system. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/nci748 |