Modeling Beam Asymmetry of the AN/UDM/1A 137Cs Calibration Source at a Radiac Calibration Facility
The AN/UDM-1A calibrator consists of a 4.44 TBq 137Cs radioisotopic source in a cylindrical lead shield. The most typical application of the unit is radiac meter calibration where predictable dose rate distributions are desired. However, vertically skewed dose rate distributions vertically from the...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 1999-02, Vol.81 (4), p.301-304 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The AN/UDM-1A calibrator consists of a 4.44 TBq 137Cs radioisotopic source in a cylindrical lead shield. The most typical application of the unit is radiac meter calibration where predictable dose rate distributions are desired. However, vertically skewed dose rate distributions vertically from the exit window centre line have been measured, and an investigation was undertaken to determine the cause. It is demonstrated that MCNP models of a well-defined problem can be developed to predict the effect of inaccuracies in the as-built configuration when compared to blueprints. The causes of the skewness are presented as well as the mitigation methodology adopted. |
---|---|