Thermoluminescence Properties of KMgF3 Doped with Pb, Cr and Ag

Thermoluminescence properties have been determined for the perovskite-like KMgF3 containing various dopant metals (Pb, Cr, Ag), in order to test the dosimetric features of these systems. The glow curves, recorded after ß irradiation, show several peaks depending on the kind of impurity, but all mate...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation protection dosimetry 1996-06, Vol.65 (1-4), p.93-96
Hauptverfasser: Kitis, G., Furetta, C., Sanipoli, C., Scacco, A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Thermoluminescence properties have been determined for the perovskite-like KMgF3 containing various dopant metals (Pb, Cr, Ag), in order to test the dosimetric features of these systems. The glow curves, recorded after ß irradiation, show several peaks depending on the kind of impurity, but all materials exhibit a typical glow peak at about 170-180 oC, which in general is well resolved and then can be considered as the dosimetric peak. Annealing of the samples at 400 oC is found to be effective for stabilisation of their sensitivity. A detailed study is carried out on low temperature pre-irradiation annealing, post-irradiation annealing, lowest detectable limit and dose response of the different systems, and a comparison of their sensitivities with that of LiF:Mg,Ti is attempted.
ISSN:0144-8420
1742-3406
DOI:10.1093/oxfordjournals.rpd.a031690