Cobas Mira™ S Endpoint Enzymatic Assay for Plasma Formate

In methanol intoxication, increased levels of the metabolite formate are associated with metabolic acidosis and an increased risk for ocular and neurological dysfunction. A simple method for plasma formate measurement by adaptation of a manual enzymatic assay to a Cobas Mira S analyzer is presented....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of analytical toxicology 2001-03, Vol.25 (2), p.77-80
Hauptverfasser: Blomme, B., Lheureux, P., Gerlo, E., Maes, V.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In methanol intoxication, increased levels of the metabolite formate are associated with metabolic acidosis and an increased risk for ocular and neurological dysfunction. A simple method for plasma formate measurement by adaptation of a manual enzymatic assay to a Cobas Mira S analyzer is presented. Six microliters of sample is incubated for 5 min with buffer containing nicotinamide-adenine dinucleotide. Fifteen microliters of a suspension of formate dehydrogenase is then added. Absorbance at 340 nm is measured every 25 s. The NADH produced when formate is oxidized is stoichiometric to the amount of formate. The method is sensitive, reproducible, and specific and has a broad measurement range. The frozen reagents are stable for at least six months, so the described method can be applied to irregular and semi-urgent requests. A recent case is reported.
ISSN:0146-4760
1945-2403