P6546Canadian Registry of Electronic Device Outcomes (CREDO): remote monitoring outcomes in the abbott battery performance alert, a multicentre cohort

Abstract Background Cardiac implantable electronic devices have been known to have lead and device malfunctions leading to advisories. Increased use of remote monitoring of devices has been suggested to allow the identification of abnormal device performance and allow early intervention. We sought t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European heart journal 2019-10, Vol.40 (Supplement_1)
Hauptverfasser: Davis, S, Thibault, B, Mangat, I, Coutu, B, Bennett, M, Philippon, F, Sandhu, R, Sterns, L, Essebad, V, Nery, P, Wells, G, Yee, R, Exner, D, Krahn, A, Parkash, R
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue Supplement_1
container_start_page
container_title European heart journal
container_volume 40
creator Davis, S
Thibault, B
Mangat, I
Coutu, B
Bennett, M
Philippon, F
Sandhu, R
Sterns, L
Essebad, V
Nery, P
Wells, G
Yee, R
Exner, D
Krahn, A
Parkash, R
description Abstract Background Cardiac implantable electronic devices have been known to have lead and device malfunctions leading to advisories. Increased use of remote monitoring of devices has been suggested to allow the identification of abnormal device performance and allow early intervention. We sought to describe the outcomes of patients with and without remote monitoring of in devices in the Abbott Premature Battery Depletion advisory with data from a Canadian registry Methods Patients with an Abbott device subject to the Battery Performance Alert Advisory from nine ICD implanting centres in Canada were included in the registry. The use of remote monitoring was identified from baseline and followup data in the registry. The primary outcome was detection of premature battery depletion and all cause mortality. Results 2679 patents were identified with a device subject to the advisory. Devices were implanted between 2010 and 2017. 1716 patients (64%) had remote monitoring at baseline with this increasing to 83.7% at followup at 12 months. Premature battery depletion occurred in 43 patients (1.6%). Discovery of premature battery depletion was detected by remote monitoring in 70% of patients. There were 492 deaths during the follow up. Mortality was higher in those without a remote monitor compared to those with a remote monitor at follow-up and remote monitor at baseline and follow-up (11.3%, 2.6% versus 6.1% respectively; p=0.0186). There were no deaths attributed to premature battery depletion Conclusion The use of remote monitoring in patients with ICD and CRT under advisory reliably detected device failure and was associated with a reduction in all-cause mortality.
doi_str_mv 10.1093/eurheartj/ehz746.1136
format Article
fullrecord <record><control><sourceid>oup_cross</sourceid><recordid>TN_cdi_crossref_primary_10_1093_eurheartj_ehz746_1136</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><oup_id>10.1093/eurheartj/ehz746.1136</oup_id><sourcerecordid>10.1093/eurheartj/ehz746.1136</sourcerecordid><originalsourceid>FETCH-LOGICAL-c876-6bbfdbf9139d1380ffd9e34f295c0608a63aba08040d32c39999926ddbd88303</originalsourceid><addsrcrecordid>eNqNkE9LwzAYh4MoOKcfQchRwc6kabPEm3TzDwwm04O3kqRv1o62GWkqzA_i57Vj6tn38l5-z3N4ELqkZEKJZLfQ-xKUD5tbKD-nCZ9QyvgRGtE0jiPJk_QYjQiVacS5eD9FZ123IYQITvkIfb3wNOGZalVRqRavYF11we-ws3hegwnetZXBM_ioDOBlH4xroMNX2Wo-W17fYQ-NC4CbYRWcr9o1dr-bqsWhBKy0diFgrUKAwbsFb51vVDvoVA0-3GCFm74Og78NHrBxpfPhHJ1YVXdw8fPH6PVh_pY9RYvl43N2v4iMmPKIa20LbSVlsqBMEGsLCSyxsUwN4UQozpRWRJCEFCw2TO4v5kWhCyEYYWOUHqzGu67zYPOtrxrldzkl-T5t_pc2P6TN92kHjhw412__iXwDpDuD0g</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>P6546Canadian Registry of Electronic Device Outcomes (CREDO): remote monitoring outcomes in the abbott battery performance alert, a multicentre cohort</title><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>Oxford University Press Journals All Titles (1996-Current)</source><source>Alma/SFX Local Collection</source><creator>Davis, S ; Thibault, B ; Mangat, I ; Coutu, B ; Bennett, M ; Philippon, F ; Sandhu, R ; Sterns, L ; Essebad, V ; Nery, P ; Wells, G ; Yee, R ; Exner, D ; Krahn, A ; Parkash, R</creator><creatorcontrib>Davis, S ; Thibault, B ; Mangat, I ; Coutu, B ; Bennett, M ; Philippon, F ; Sandhu, R ; Sterns, L ; Essebad, V ; Nery, P ; Wells, G ; Yee, R ; Exner, D ; Krahn, A ; Parkash, R</creatorcontrib><description>Abstract Background Cardiac implantable electronic devices have been known to have lead and device malfunctions leading to advisories. Increased use of remote monitoring of devices has been suggested to allow the identification of abnormal device performance and allow early intervention. We sought to describe the outcomes of patients with and without remote monitoring of in devices in the Abbott Premature Battery Depletion advisory with data from a Canadian registry Methods Patients with an Abbott device subject to the Battery Performance Alert Advisory from nine ICD implanting centres in Canada were included in the registry. The use of remote monitoring was identified from baseline and followup data in the registry. The primary outcome was detection of premature battery depletion and all cause mortality. Results 2679 patents were identified with a device subject to the advisory. Devices were implanted between 2010 and 2017. 1716 patients (64%) had remote monitoring at baseline with this increasing to 83.7% at followup at 12 months. Premature battery depletion occurred in 43 patients (1.6%). Discovery of premature battery depletion was detected by remote monitoring in 70% of patients. There were 492 deaths during the follow up. Mortality was higher in those without a remote monitor compared to those with a remote monitor at follow-up and remote monitor at baseline and follow-up (11.3%, 2.6% versus 6.1% respectively; p=0.0186). There were no deaths attributed to premature battery depletion Conclusion The use of remote monitoring in patients with ICD and CRT under advisory reliably detected device failure and was associated with a reduction in all-cause mortality.</description><identifier>ISSN: 0195-668X</identifier><identifier>EISSN: 1522-9645</identifier><identifier>DOI: 10.1093/eurheartj/ehz746.1136</identifier><language>eng</language><publisher>Oxford University Press</publisher><ispartof>European heart journal, 2019-10, Vol.40 (Supplement_1)</ispartof><rights>Published on behalf of the European Society of Cardiology. All rights reserved. © The Author(s) 2019. For permissions, please email: journals.permissions@oup.com. 2019</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,1584,27924,27925</link.rule.ids></links><search><creatorcontrib>Davis, S</creatorcontrib><creatorcontrib>Thibault, B</creatorcontrib><creatorcontrib>Mangat, I</creatorcontrib><creatorcontrib>Coutu, B</creatorcontrib><creatorcontrib>Bennett, M</creatorcontrib><creatorcontrib>Philippon, F</creatorcontrib><creatorcontrib>Sandhu, R</creatorcontrib><creatorcontrib>Sterns, L</creatorcontrib><creatorcontrib>Essebad, V</creatorcontrib><creatorcontrib>Nery, P</creatorcontrib><creatorcontrib>Wells, G</creatorcontrib><creatorcontrib>Yee, R</creatorcontrib><creatorcontrib>Exner, D</creatorcontrib><creatorcontrib>Krahn, A</creatorcontrib><creatorcontrib>Parkash, R</creatorcontrib><title>P6546Canadian Registry of Electronic Device Outcomes (CREDO): remote monitoring outcomes in the abbott battery performance alert, a multicentre cohort</title><title>European heart journal</title><description>Abstract Background Cardiac implantable electronic devices have been known to have lead and device malfunctions leading to advisories. Increased use of remote monitoring of devices has been suggested to allow the identification of abnormal device performance and allow early intervention. We sought to describe the outcomes of patients with and without remote monitoring of in devices in the Abbott Premature Battery Depletion advisory with data from a Canadian registry Methods Patients with an Abbott device subject to the Battery Performance Alert Advisory from nine ICD implanting centres in Canada were included in the registry. The use of remote monitoring was identified from baseline and followup data in the registry. The primary outcome was detection of premature battery depletion and all cause mortality. Results 2679 patents were identified with a device subject to the advisory. Devices were implanted between 2010 and 2017. 1716 patients (64%) had remote monitoring at baseline with this increasing to 83.7% at followup at 12 months. Premature battery depletion occurred in 43 patients (1.6%). Discovery of premature battery depletion was detected by remote monitoring in 70% of patients. There were 492 deaths during the follow up. Mortality was higher in those without a remote monitor compared to those with a remote monitor at follow-up and remote monitor at baseline and follow-up (11.3%, 2.6% versus 6.1% respectively; p=0.0186). There were no deaths attributed to premature battery depletion Conclusion The use of remote monitoring in patients with ICD and CRT under advisory reliably detected device failure and was associated with a reduction in all-cause mortality.</description><issn>0195-668X</issn><issn>1522-9645</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><recordid>eNqNkE9LwzAYh4MoOKcfQchRwc6kabPEm3TzDwwm04O3kqRv1o62GWkqzA_i57Vj6tn38l5-z3N4ELqkZEKJZLfQ-xKUD5tbKD-nCZ9QyvgRGtE0jiPJk_QYjQiVacS5eD9FZ123IYQITvkIfb3wNOGZalVRqRavYF11we-ws3hegwnetZXBM_ioDOBlH4xroMNX2Wo-W17fYQ-NC4CbYRWcr9o1dr-bqsWhBKy0diFgrUKAwbsFb51vVDvoVA0-3GCFm74Og78NHrBxpfPhHJ1YVXdw8fPH6PVh_pY9RYvl43N2v4iMmPKIa20LbSVlsqBMEGsLCSyxsUwN4UQozpRWRJCEFCw2TO4v5kWhCyEYYWOUHqzGu67zYPOtrxrldzkl-T5t_pc2P6TN92kHjhw412__iXwDpDuD0g</recordid><startdate>20191001</startdate><enddate>20191001</enddate><creator>Davis, S</creator><creator>Thibault, B</creator><creator>Mangat, I</creator><creator>Coutu, B</creator><creator>Bennett, M</creator><creator>Philippon, F</creator><creator>Sandhu, R</creator><creator>Sterns, L</creator><creator>Essebad, V</creator><creator>Nery, P</creator><creator>Wells, G</creator><creator>Yee, R</creator><creator>Exner, D</creator><creator>Krahn, A</creator><creator>Parkash, R</creator><general>Oxford University Press</general><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>20191001</creationdate><title>P6546Canadian Registry of Electronic Device Outcomes (CREDO): remote monitoring outcomes in the abbott battery performance alert, a multicentre cohort</title><author>Davis, S ; Thibault, B ; Mangat, I ; Coutu, B ; Bennett, M ; Philippon, F ; Sandhu, R ; Sterns, L ; Essebad, V ; Nery, P ; Wells, G ; Yee, R ; Exner, D ; Krahn, A ; Parkash, R</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c876-6bbfdbf9139d1380ffd9e34f295c0608a63aba08040d32c39999926ddbd88303</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Davis, S</creatorcontrib><creatorcontrib>Thibault, B</creatorcontrib><creatorcontrib>Mangat, I</creatorcontrib><creatorcontrib>Coutu, B</creatorcontrib><creatorcontrib>Bennett, M</creatorcontrib><creatorcontrib>Philippon, F</creatorcontrib><creatorcontrib>Sandhu, R</creatorcontrib><creatorcontrib>Sterns, L</creatorcontrib><creatorcontrib>Essebad, V</creatorcontrib><creatorcontrib>Nery, P</creatorcontrib><creatorcontrib>Wells, G</creatorcontrib><creatorcontrib>Yee, R</creatorcontrib><creatorcontrib>Exner, D</creatorcontrib><creatorcontrib>Krahn, A</creatorcontrib><creatorcontrib>Parkash, R</creatorcontrib><collection>CrossRef</collection><jtitle>European heart journal</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Davis, S</au><au>Thibault, B</au><au>Mangat, I</au><au>Coutu, B</au><au>Bennett, M</au><au>Philippon, F</au><au>Sandhu, R</au><au>Sterns, L</au><au>Essebad, V</au><au>Nery, P</au><au>Wells, G</au><au>Yee, R</au><au>Exner, D</au><au>Krahn, A</au><au>Parkash, R</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>P6546Canadian Registry of Electronic Device Outcomes (CREDO): remote monitoring outcomes in the abbott battery performance alert, a multicentre cohort</atitle><jtitle>European heart journal</jtitle><date>2019-10-01</date><risdate>2019</risdate><volume>40</volume><issue>Supplement_1</issue><issn>0195-668X</issn><eissn>1522-9645</eissn><abstract>Abstract Background Cardiac implantable electronic devices have been known to have lead and device malfunctions leading to advisories. Increased use of remote monitoring of devices has been suggested to allow the identification of abnormal device performance and allow early intervention. We sought to describe the outcomes of patients with and without remote monitoring of in devices in the Abbott Premature Battery Depletion advisory with data from a Canadian registry Methods Patients with an Abbott device subject to the Battery Performance Alert Advisory from nine ICD implanting centres in Canada were included in the registry. The use of remote monitoring was identified from baseline and followup data in the registry. The primary outcome was detection of premature battery depletion and all cause mortality. Results 2679 patents were identified with a device subject to the advisory. Devices were implanted between 2010 and 2017. 1716 patients (64%) had remote monitoring at baseline with this increasing to 83.7% at followup at 12 months. Premature battery depletion occurred in 43 patients (1.6%). Discovery of premature battery depletion was detected by remote monitoring in 70% of patients. There were 492 deaths during the follow up. Mortality was higher in those without a remote monitor compared to those with a remote monitor at follow-up and remote monitor at baseline and follow-up (11.3%, 2.6% versus 6.1% respectively; p=0.0186). There were no deaths attributed to premature battery depletion Conclusion The use of remote monitoring in patients with ICD and CRT under advisory reliably detected device failure and was associated with a reduction in all-cause mortality.</abstract><pub>Oxford University Press</pub><doi>10.1093/eurheartj/ehz746.1136</doi></addata></record>
fulltext fulltext
identifier ISSN: 0195-668X
ispartof European heart journal, 2019-10, Vol.40 (Supplement_1)
issn 0195-668X
1522-9645
language eng
recordid cdi_crossref_primary_10_1093_eurheartj_ehz746_1136
source Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; Oxford University Press Journals All Titles (1996-Current); Alma/SFX Local Collection
title P6546Canadian Registry of Electronic Device Outcomes (CREDO): remote monitoring outcomes in the abbott battery performance alert, a multicentre cohort
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-30T20%3A42%3A02IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-oup_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=P6546Canadian%20Registry%20of%20Electronic%20Device%20Outcomes%20(CREDO):%20remote%20monitoring%20outcomes%20in%20the%20abbott%20battery%20performance%20alert,%20a%20multicentre%20cohort&rft.jtitle=European%20heart%20journal&rft.au=Davis,%20S&rft.date=2019-10-01&rft.volume=40&rft.issue=Supplement_1&rft.issn=0195-668X&rft.eissn=1522-9645&rft_id=info:doi/10.1093/eurheartj/ehz746.1136&rft_dat=%3Coup_cross%3E10.1093/eurheartj/ehz746.1136%3C/oup_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rft_oup_id=10.1093/eurheartj/ehz746.1136&rfr_iscdi=true