P6546Canadian Registry of Electronic Device Outcomes (CREDO): remote monitoring outcomes in the abbott battery performance alert, a multicentre cohort
Abstract Background Cardiac implantable electronic devices have been known to have lead and device malfunctions leading to advisories. Increased use of remote monitoring of devices has been suggested to allow the identification of abnormal device performance and allow early intervention. We sought t...
Gespeichert in:
Veröffentlicht in: | European heart journal 2019-10, Vol.40 (Supplement_1) |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Background
Cardiac implantable electronic devices have been known to have lead and device malfunctions leading to advisories. Increased use of remote monitoring of devices has been suggested to allow the identification of abnormal device performance and allow early intervention. We sought to describe the outcomes of patients with and without remote monitoring of in devices in the Abbott Premature Battery Depletion advisory with data from a Canadian registry
Methods
Patients with an Abbott device subject to the Battery Performance Alert Advisory from nine ICD implanting centres in Canada were included in the registry. The use of remote monitoring was identified from baseline and followup data in the registry. The primary outcome was detection of premature battery depletion and all cause mortality.
Results
2679 patents were identified with a device subject to the advisory. Devices were implanted between 2010 and 2017. 1716 patients (64%) had remote monitoring at baseline with this increasing to 83.7% at followup at 12 months. Premature battery depletion occurred in 43 patients (1.6%). Discovery of premature battery depletion was detected by remote monitoring in 70% of patients. There were 492 deaths during the follow up. Mortality was higher in those without a remote monitor compared to those with a remote monitor at follow-up and remote monitor at baseline and follow-up (11.3%, 2.6% versus 6.1% respectively; p=0.0186). There were no deaths attributed to premature battery depletion
Conclusion
The use of remote monitoring in patients with ICD and CRT under advisory reliably detected device failure and was associated with a reduction in all-cause mortality. |
---|---|
ISSN: | 0195-668X 1522-9645 |
DOI: | 10.1093/eurheartj/ehz746.1136 |